1700°C 12-zone Split Tube Furnace with a Ø100 mm × 1800 mm L Alumina Tube - OTF1700X12Z

1700°C 12-zone Split Tube Furnace with a Ø100 mm × 1800 mm L Alumina Tube - OTF1700X12Z

RM0.00 MYR
Sale price  RM0.00 MYR Regular price 
Skip to product information
1700°C 12-zone Split Tube Furnace with a Ø100 mm × 1800 mm L Alumina Tube - OTF1700X12Z

1700°C 12-zone Split Tube Furnace with a Ø100 mm × 1800 mm L Alumina Tube - OTF1700X12Z

SKU: otf1700x12z

Price: RFQ

Description

OTF-1700X-12Z is a 12-zone split tube furnace up to 1700°C It is a perfect tool to precisely control and build variable temperature gradients along the entire heating area to simulate reactive kinetics and mechanism in a lab scale.

Features

  • Temperature up to 1700°C
  • Heating length 1400 mm 1
  • 2 heating zones, 12 individual PID temperature controllers.
  • Ø100 mm × 1800 mm alumina tube

SPECIFICATIONS

Power
  • 480VAC, 3phase, 50/60 Hz 
  • 60 KVA
  • Power cable is included for hard wiring.
Furnace Case
  • Double-layer mild steel case with cooling fans.
  • Surface temperature < 70°C

Heating Zones
Thermocouples
  • Type S thermocouple for Zone 1-5, 11-12
  • Type B thermocouple for Zone 6-10

Temperature Controller
  • 12 PID programmable temperature controllers
  • 50 temperature segments
  • Overheating and broken thermocouple interlock/alarm.
  • ±1°C control accuracy
  • 0.3°C temperature resolution
  • RS485 serial port with LabVIEW-based software
Processing Tube

         

Flanges
  • Stainless steel 304 water-cooled flange
  • 1/4" tube gas inlet
  • -0.1 - 0.5 MPa diaphragm pressure gauge
  • KF25 vacuum port
  • DN100CF or ISO100K service port
  • (Optional) Port aligner
Chilled Water
  • 5 - 35°C, 10 L/min, 10 psi chilled water is required.
  • MTI recommend using KJ5000 water chiller  
Pressure
  • < 3 psig
  • <3e-2 torr with a vacuum pump
Vacuum (Optional)
Compliance
  • CE Certified
  • NRTL or CSA certification is available upon request at the extra cost
Net Weight
  • 660 kg
Application Notes
  • 3-day 2-engineer on-site installation is required.