Skip to product information
150L Programmable Fast Thermal Test Chamber (-60C to +150C, UN38.3.4.2) - MSK-TE906-150L

150L Programmable Fast Thermal Test Chamber (-60C to +150C, UN38.3.4.2) - MSK-TE906-150L

SKU: 150l-programmable-fast-thermal-test-chamber-60c-to-150c-un38-3-4-2-msk-te906-150l

Price: RFQ

Specifications

MSK-TE906 is a 150-liter programmable thermal test chamber, capable of studying the performance of rechargeable battery at rapidly changing the temperature in the range from -60°C to 150°C. The chamber contains a built-in air blower which enables to achieve a uniform temperature field. The test temperature is consciously monitored by a built-in thermocouple and can be precisely controlled by the integrated PLC system. The thermal test system is built based on the thermal test requirement (38.3.4.2) of lithium battery testing requirements of the UN Manual of tests and criteria. An observation window is mounted on the front panel, allowing to monitor the real-time status of rechargeable batteries. The draining system could easily collect unexpected liquids. Please click here to review the UN38.3 Li-Ion Battery Transportation Safety Testing Requirements SPECIFICATIONS Test Chamber Features SS304 stainless steel chamber with foam polyurethane and fiberglass as the thermal insulator Chamber dimensions: 500 × 600 × 500 mm (L × D × H) Window: 270 mm W x 370 mm, vacuum-sealed Capacity: 150 Liters Vent Port: Vent port (310x412mm) built in the back for releasing pressure or smoking in the case battery Port: Two sealed ports (100 mm Dia.) are on the left for wire feed-through Optional: customized chamber size is available upon request Control Panel 6" touch-screen control panel to set up the program, and display temperature profile 1200 segments programmable, and each program can be repeated up to 999 times Cooling & Heating System Dual refrigeration compressor made in France (1.5 HP) Refrigerant: R404A, environmentally friendly Stainless heater tube Built-in beep alarm for over-heated and over-current protection Working Temperature: programmable from -60ºC to 150ºC Temperature accuracy: +/- 1ºC Max. Heating rate: 5 ºC/min Max. Cooling rate: 3~4 ºC/min Calibration You may order NIST certified temperature monitor to calibrate the temperature of the inside chamber at an extra cost (click the pic to order) Power Voltage: 208 - 230VAC (3 phase, 50/60 Hz), 380 VAC or 460 VAC with a 10kW transformer Maximum current: 50 A Power: 8 kW Compliance CE certified Built based on TUV, CSA, and UL, and ready to pass certificated at an extra cost Optional Another standard chamber size is available at 225 Liters (Chamber 500 x 750 x 600 mm, Overall 1170 x 1840 x 1030 mm) 408 Liters (Chamber 600 x 850 x 800 mm, Overall 1100 x 1900 x 1900 mm) 800 Liters (Chamber 1000 x 1000 x 800 mm, Overall 1570 x 2070 x 1320, mm) Application Notes Please Click Here for Details on UN38.3 Li-Ion Battery Testing Requirements Other equipments for testing of IEC 62133 or UN38.3 Standards can be found at MTI

Test Chamber Features
 
  • SS304 stainless steel chamber with foam polyurethane and fiberglass as the thermal insulator
  • Chamber dimensions: 500 × 600 × 500 mm (L × D × H)
  • Window: 270 mm W x 370 mm, vacuum-sealed
  • Capacity: 150 Liters
  • Vent Port: Vent port (310x412mm) built in the back for releasing pressure or smoking in the case battery
  • Port: Two sealed ports (100 mm Dia.) are on the left for wire feed-through
  • Optional: customized chamber size is available upon request
Control Panel
  • 6" touch-screen control panel to set up the program, and display temperature profile
  • 1200 segments programmable, and each program can be repeated up to 999 times



Cooling & Heating System
  • Dual refrigeration compressor made in France (1.5 HP)
  • Refrigerant: R404A, environmentally friendly
  • Stainless heater tube
  • Built-in beep alarm for over-heated and over-current protection
  • Working Temperature: programmable from -60ºC to 150ºC
  • Temperature accuracy: +/- 1ºC
  • Max. Heating rate: 5 ºC/min
  • Max. Cooling rate: 3~4 ºC/min

Calibration
  • You may order NIST certified temperature monitor to calibrate the temperature of the inside chamber at an extra cost
    (click the pic to order)

Power
  • Voltage: 208 - 230VAC (3 phase, 50/60 Hz), 380 VAC or 460 VAC with a 10kW transformer
  • Maximum current: 50 A
  • Power: 8 kW
Compliance
  •  CE certified
  •  Built based on TUV, CSA, and UL, and ready to pass certificated at an extra cost


Optional
  • Another standard chamber size is available at
    • 225 Liters (Chamber 500 x 750 x 600 mm, Overall 1170 x 1840 x 1030 mm)
    • 408 Liters (Chamber 600 x 850 x 800 mm, Overall 1100 x 1900 x 1900 mm)
    • 800 Liters (Chamber 1000 x 1000 x 800 mm, Overall 1570 x 2070 x 1320, mm)
Application Notes

You might also like...